STUDY OBJECTIVES AND METHODS
The aim of this study was to examine the effect of insertion load on implant primary stability by evaluating the insertion torque and insertion time in various implant designs. Four implant designs were tested, including one cylindrical implant standard (S), two hybrid implants tapered effect (TE) and bone level (BL), and one conical implant bone level tapered (BLT). Polyurethane bone models of the maxillary posterior region were used. Insertion torque value (ITV) and insertion time, defined as the duration from implant placement initiation to platform alignment, were recorded under two load conditions, the minimum load and a load of 5.0 newton (N). A torque meter was used to capture torque-time curves, and the mean and standard deviation of ITV were calculated. Data were analyzed using a paired t-test (P < 0.05).
RESULTS
- The minimum insertion load varied by design: implant S required 2.5 N, implants TE and BL each required 2.0 N, and implant BLT required 1.0 N.
- At minimum load, insertion torque was 8.68 N cm for implant S, 6.64 N cm for implant TE, 12.29 N cm for implant BL, and 29.52 N cm for implant BLT.
- Under 5.0 N, the values were 8.12, 7.82, 14.89, and 30.53 N cm, respectively. Insertion time decreased by up to 12.52 % from 1.0 N to 5.0 N, with significant differences in implant BLT.
CONCLUSIONS
Hybrid implants are more sensitive to load variations. Optimizing the insertion load based on implant design can enhance clinical outcomes. The insertion load is a critical but often overlooked factor in primary implant stability.
Adapted from Po-Yuan Hsueh et al., J Dent Sci. 2025 Jul;20(3):1861-1868, for more info about this publication, click HERE.
This review is part of the Straumann "Scientific Highlights Newsletter 4/25".
More studies:
- Five Year Clinical, Radiographic and Soft Tissue Profilometric Outcomes at Two Narrow-Diameter Implants to Replace Missing Maxillary Lateral Incisors
- Osseointegration in the Absence of Primary Stability: An Experimental Preclinical Mandibular Minipig Overpreparation In Vivo Model
- Radiographic Peri-Implant Bone Changes in Osteoporotic Women Treated With a Ti-Zr, Bone Level Tapered Implant With a Hydrophilic Surface: A 12-Month Prospective Case-Series
- Role of Keratinized Mucosa on the Risk of Peri-Implant Diseases and Soft Tissue Dehiscence in the Posterior Mandible-A 20-Year Prospective Cohort Study
- Analysis of bone dimensional stability after two-stage maxillary sinus floor augmentation with autogenous bone versus bovine bone mineral combined with autogenous bone chips: results from a 1-year multicenter split-mouth randomized controlled trial
- Early Buccal Bone Resorption in Areas With or Without Keratinized Tissue and Different Mucosal Thickness at Implants in Healed Sites: An Experimental Animal Study